This procedure is provided for the
purpose of determining the percentage of test material that is volatilized
and driven off under the conditions specified. The procedure, as generally
applied, is nondestructive to the substance under test; however, the substance
may be converted to another form such as an anhydride.
Perform the test on finely powdered
material, and break up lumps, if necessary, with the aid of a mortar and
pestle before weighing the specimen. Weigh the specimen to be tested without
further treatment, unless a preliminary drying at a lower temperature,
or other special pretreatment, is specified in the individual monograph.
Unless other equipment is
10/L, in which L is the limit (or the mean
value of the limits) for Loss on ignition, in percentage. Ignite the loaded
uncovered crucible, and cover at the temperature ( ± 25º) and
for the period of time designated in the individual monograph. Ignite for
successive 1 hour periods where ignition to constant weight is indicated.
Upon completion of each ignition, cover the crucible, and allow it to cool
in a desiccator lo room temperature before weighing.
|